SKC Real Time Monitor
Set the pump to 10 L/min. If the calibrator displays 9.7 L/min, click the bu on
in Flow Calibrate until the calibrator displays 10 L/min. If the calibrator displays
10.5 L/min, click the button in Flow Calibrate until the calibrator displays
10 L/min. Repeat calibration a er sampling to verify ow.
Comm Checking Buttons
The Comm Checking bu ons (Figure 9) turn the communication checking function
on or o . Comm Checking monitors the interface cable connection between the PC
and the pump. The default value is On. If the interface cable becomes detached, an
error message displays (Figure 10). Reconnect the pump and click on Retry. If the
O bu on is selected in Comm Checking, the pump’s real time information will
not be updated.
If programming more than one pump, turn Comm Checking o by clicking the
O bu on. Turning Comm Checking o when programming multiple pumps will
eliminate the error message that displays each time the pump is disconnected.
When Comm Checking is turned o , the pump will enter Sleep mode ve minutes
a er the last interaction between the computer and the pump.
Figure 9.
Comm Checking Button
Figure 10.
Error Message