Reports
Figure 42.
Calibration Info Window with Con rm Validation Code
Validating CalChek Full Calibration Data
To con rm printed calibration data, open the DataTrac Pump Manager window
and click on the View menu. Choose Calibration Info (Figure 41). Click the Print
Report bu on. Go to the Tools menu and choose Con rm Validation Code (Figure
42). Enter the calibration date shown on the printed report, enter each actual ow,
and then enter the validation code. Click on the Check Validation Code bu on.
The box to the right of the bu on will display red and “invalid” if the data has
been entered incorrectly or tampered with. A green bar with the word “valid” will
display if data entered is valid.
Clearing the history will not clear full calibration data. This data can only be
cleared by performing another full calibration.
When entering data to con rm the validation number, enter the date in the
following format: mmm dd, yyyy (e.g., Aug 18, 2010).